Type-52 [for solvent gases (%lel) – COSA Xentaur SF6 Purity Pro User Manual
Page 26
26
TYPE-52 [FOR SOLVENT GASES (%LEL)]
Selectable Measuring Gas and Range
Measuring Gas
(Display Indication)
Measuring
Range
(Minimum
Digit)
AIR CAL
Value
Drift
Amount
Spec. No.
1
Toluene in AIR
(C7H8 in AIR)
0~100 %LEL
(0.2)
0.0 %LEL 1.4 %LEL
SPE- 1434
2
Methyl ethyl ketone
in AIR
(MEK in AIR)
0~100 %LEL
(0.2)
0.0 %LEL 1.5 %LEL
SPE- 1435
3
Ethyl acetate in AIR
(EtAc in AIR)
0~100 %LEL
(0.2)
0.0 %LEL 1.1 %LEL
SPE- 1436
4
Xylene in AIR
(C8H10 in AIR)
0~100 %LEL
(0.2)
0.0 %LEL 1.4 %LEL
SPE- 1437
5
Isopropyl alcohol in
AIR
(IPA in AIR)
0~100 %LEL
(0.2)
0.0 %LEL 1.7 %LEL
SPE- 1438
6
Methyl isobutyl
ketone in AIR
(MIBK in AIR)
0~100 %LEL
(0.2)
0.0 %LEL 1.3 %LEL
SPE- 1439
7
Methanol in AIR
(CH3OH in AIR)
0~100 %LEL
(0.2)
0.0 %LEL 2.0 %LEL
SPE- 1447
8
-----------------
-----------------
-------------- ----------------- -----------------
Interferometer Type: Type 120MM
Indication Accuracy: ± (Indication Value – AIR CAL Value) x 0.03 ± Drift Amount
AIR CAL Value: Concentration indicated on the display when AIR CAL key is pressed.
Drift Amount: The amount of indication drifting when temperature changes ±10ºC.
Special Caution:
4) This instrument is not explosion proof.
5) This instrument measures combustible gas around lower explosive limits in this
specification. Take enough sample for the measurement.
6) Measuring gases in this specification are not extensively vaporized. The gas
concentration may not reach a higher concentration in themeasuring range depending
on a condition of temperature and pressure.