beautypg.com

Trouble shooting, Symptoms causes solutions – COSA Xentaur SF6 Purity Pro User Manual

Page 18

background image

18

6. TROUBLE SHOOTING

The trouble shootings mentioned below do not cover all problems, however those that occur most

frequently are indicated to help your research for cause and solution.

Symptoms

Causes

Solutions

Power does not turn ON.

No batteries; AC adaptor may

not be connected.

Polarity of the batteries is

wrong.

The POWER key has not been

pressed long enough.

Insert batteries; connect AC

adaptor.

Place batteries in the correct

direction.

Continue to press POWER key

until the buzzer beeps.

Reading is not correct.

AIR CAL is not functioning

correctly.

The incorrect “Measuring Gas”

or “Base Gas” has been

selected.

Interference gas is included in

the sample gas.

A sudden pressure change

occurs at GAS IN, GAS OUT, or

REF. OUT.

The pressure at GAS IN, GAS

OUT or REF. OUT is not at

atmospheric pressure.

When the instrument samples

enough fresh air, stop the

pump. Then press the AIR CAL

key.

Select the correct “Measuring

Gas” and “Base Gas.”

The target gas that contains

interference gas cannot be

measured.

Perform AIR CAL again.

Set GAS IN, GAS OUT, and REF.

OUT under the atmospheric

pressure.

After performing AIR CAL, the

reading goes up even though

no gas has been fed.

Molecular attached to the

chamber or tubing inside

begins to detach. The

concentration in the chamber

becomes high.

Let the instrument sample

some fresh air and confirm if it

shows the AIR CAL value.