Type-55 [for anaesthetic gases – COSA Xentaur SF6 Purity Pro User Manual
Page 29
29
TYPE-55 [FOR ANAESTHETIC GASES]
Selectable Measuring Gas and Range
Measuring Gas
(Display Indication)
Measuring
Range
(Minimum
Digit)
AIR CAL
Value
Drift
Amount
Spec. No.
1
Halothane in O2
(HALOTHANE in O2)
0~6 vol%
(0.01)
1.57 vol% 0.09 vol% SPE- 1401
2
Isoflurane in O2
(ISOFLURANE in 02)
0~8 vol%
(0.01)
1.65 vol% 0.09 vol% SPE- 1402
3
Sevoflurane in O2
(SEVOFLURANE in
O2)
0~10 vol%
(0.01)
1.70 vol% 0.09 vol% SPE- 1403
4
Desflurane in O2
(DESFLURANE in O2)
0~20 vol%
(0.02)
2.10 vol% 0.12 vol% SPE- 1404
5
Halothane in AIR
(HALOTHANE in AIR)
0~6 vol%
(0.01)
0.00 vol% 0.09 vol% SPE- 1397
6
Isoflurane in AIR
(ISOFLURANE in AIR)
0~8 vol%
(0.01)
0.00 vol% 0.09 vol% SPE- 1398
7
Sevoflurane in AIR
(SEVOFLURANE in
AIR)
0~10 vol%
(0.01)
0.00 vol% 0.10 vol% SPE- 1399
8
Desflurane in AIR
(DESFLURANE in AIR)
0~20 vol%
(0.02)
0.00 vol% 0.12 vol% SPE- 1400
Interferometer Type: Type 30MM
Indication Accuracy: ± (Indication Value – AIR CAL Value) x 0.03 ± Drift Amount
AIR CAL Value: Concentration indicated on the display when AIR CAL key is pressed.
Drift Amount: The amount of indication drifting when temperature changes ±10ºC.
Special Caution:
1) Measuring gases in this specification are not extensively vaporized. The gas
concentration may not reach a higher concentration in the measuring range depending
on the conditions of temperature and pressure.