Type-55 [for anaesthetic gases – COSA Xentaur SF6 Purity Pro User Manual

Page 29

background image

29

TYPE-55 [FOR ANAESTHETIC GASES]

Selectable Measuring Gas and Range

Measuring Gas

(Display Indication)

Measuring

Range

(Minimum

Digit)

AIR CAL

Value

Drift

Amount

Spec. No.

1

Halothane in O2

(HALOTHANE in O2)

0~6 vol%

(0.01)

1.57 vol% 0.09 vol% SPE- 1401

2

Isoflurane in O2

(ISOFLURANE in 02)

0~8 vol%

(0.01)

1.65 vol% 0.09 vol% SPE- 1402

3

Sevoflurane in O2

(SEVOFLURANE in

O2)

0~10 vol%

(0.01)

1.70 vol% 0.09 vol% SPE- 1403

4

Desflurane in O2

(DESFLURANE in O2)

0~20 vol%

(0.02)

2.10 vol% 0.12 vol% SPE- 1404

5

Halothane in AIR

(HALOTHANE in AIR)

0~6 vol%

(0.01)

0.00 vol% 0.09 vol% SPE- 1397

6

Isoflurane in AIR

(ISOFLURANE in AIR)

0~8 vol%

(0.01)

0.00 vol% 0.09 vol% SPE- 1398

7

Sevoflurane in AIR

(SEVOFLURANE in

AIR)

0~10 vol%

(0.01)

0.00 vol% 0.10 vol% SPE- 1399

8

Desflurane in AIR

(DESFLURANE in AIR)

0~20 vol%

(0.02)

0.00 vol% 0.12 vol% SPE- 1400

Interferometer Type: Type 30MM

Indication Accuracy: ± (Indication Value – AIR CAL Value) x 0.03 ± Drift Amount

AIR CAL Value: Concentration indicated on the display when AIR CAL key is pressed.

Drift Amount: The amount of indication drifting when temperature changes ±10ºC.

Special Caution:

1) Measuring gases in this specification are not extensively vaporized. The gas

concentration may not reach a higher concentration in the measuring range depending

on the conditions of temperature and pressure.