Type-54 [for gas purity – COSA Xentaur SF6 Purity Pro User Manual
Page 28
28
TYPE-54 [FOR GAS PURITY]
Selectable Measuring Gas and Range
Measuring Gas
(Display Indication)
Measuring
Range
(Minimum
Digit)
AIR CAL
Value
Drift
Amount
Spec. No.
1
Helium in AIR
(He in AIR)
0~100 vol%
(0.05)
0.00 vol% 0.46 vol% SPE- 1411
2
Neon in AIR
(Ne in AIR)
0~100 vol%
(0.1)
0.0 vol%
0.5 vol%
SPE- 1442
3
Hydrogen in AIR
(H2 in AIR)
0~100 vol%
(0.1)
0.0 vol%
0.8 vol%
SPE- 1412
4
Methane in AIR
(CH4 in AIR)
0~100 vol%
(0.1)
0.0 vol%
0.7 vol%
SPE- 1443
5 Carbon dioxide in AIR
(CO2 in AIR)
0~100 vol%
(0.1)
0.0 vol%
0.8 vol%
SPE- 1413
6
Ammonia in N2
(NH3 in N2)
0~100 vol%
(0.2)
-6.6 vol%
1.4 vol%
SPE- 1418
7
Hydrogen in N2
(H2 in N2)
0~100 vol%
(0.1)
3.5 vol%
0.7 vol%
SPE- 1420
8
Hydrogen in CO2
(H2 in CO2)
0~100 vol%
(0.05)
50.20
vol%
0.38 vol% SPE- 1444
Interferometer Type: Type 30MM
Indication Accuracy: ± (Indication Value – AIR CAL Value) x 0.03 ± Drift Amount
AIR CAL Value: Concentration indicated on the display when AIR CAL key is pressed.
Drift Amount: The amount of indication drifting when temperature changes ±10ºC.
Special Caution:
1) This instrument is not explosion proof. This instrument measures combustible gas
around lower explosive limits in this specification. Take enough sample for the
measurement.
2) This instrument measures extremely toxic gases. It exhausts sample gas from GAS
OUT, and sample gas may leak from internal piping. Take enough sample for the
measurement.