Series 72 manual page 21 – COSA Xentaur Hygrocontrol Type 72 User Manual
Page 25
Series 72 Manual page 21
PSF = Saturated Vapor Pressure over Chilled Water
PSE = Saturated Vapor Pressure over Ice
M
= Rel. Humidity displayed by HYGROMESS Instruments.
11.7 Measurements under Pressure or Vaccuum
All probes can be used at pressures of 0,9 to 1,3 bar. Only special probes
with a glas feed-through (which predicts SS - 316 tube material) may be
used under pressures of 0,03 to 30 bar. As rel. Humidity is directly
proportional to the pressure (f.E.: increase of pressure by factor 2 also
increases rel. Humidity by factor 2, when temperature and content of water
remain constant!) you shoud take care of a place with constant pressure
values, when mounting the probe. Be aware that pressure may change,
when diameters of tubes change - and air streams may be turbulent at
these places.
Table 4: Gas Concentrations with Humidity Error < 2.5 %RH
MAXIMUM CONCENTRATION AT
MAK-
CONCENTRATIO
N
ACCORDING TO
continuous load
50 % load
EXPL.-
LIMIT
CHEMICAL
ppm
g/m
3
ppm
g/m
3
ppm
g/m
3
g/m
3
Ethanol (Alkohol)
1000 1,9
3500 6
7000 12
57
Isopropanol
400 0,98
4800 12
10000 25
67
Toluol + Xylol
100 0,38
1300 5 3000 12
53
Gasoline (pure)
300 1,1-1,4
150
200
51
Gasoline super (Automobil)
100
150
32
Ethylenglycol
100 0,26
1200 3 1200 3
80
Acetone
1000 2,4
3300 8
6500 16
56
Ethylacetat
400 1,4
4000
15
8000 30
79
Acetic Acid
10 0,03
800 2 1200 3
107
Ammoniumhydroxid
25
0,02
5500
4
11000
8 -
Chlorine Acid (HCl)
5
0,01
300
0,5
500
0,75 -
Sulfurhydrogen (H
2
S)
10
0,01
350
0,5
700
1 -
Sulfurdioxid (SO
2
)
5
0,01
5
0,01
5
0,01 -
1) MAK = max. Concentration at Labor (at 1 bar and 20 °C)
2) ppm = g/m
3
x 24,04/MOL
MOL = Molweight of Material