Data format UPP 35

9 Data format UPP (Universal Pyrometer

Protocol)

Via interface and suitable communication software or via Test function of the InfraWin

software commands can be exchanged directly with the pyrometer.
The data exchange occurs in ASCII format with the following transmission parameters:
The data format is: 8 data bits, 1 stop bit, even parity (8,1,e).
The device responds to the entry of a command with: output (e.g. the measuring value) + CR
(Carriage Return, ASCII 13), to pure entry commands with ok + CR.
Every command starts with the 2-digit device address AA (e.g. "00"). This is followed by 2 small
command letters (e.g. "em" for level of emissivity ), finished with CR.
This is followed, if necessary for that command, by the ASCII parameter "X". If this parameter

"X" is omitted, then the device resets with the current parameter.

A ? after the small command letters answers with the respective settings (only at setting

commands, not at enquiry commands).

Example:

Entry:

“00em“

+ CR

The

emissivity

setting

( ) of the device with the address 00 is returned.

Answer:

“0970“

+ CR means Emissivity = 0.97 or 97.0%

Description

Command

Parameters

Reading
temperature

value:

AAms

Output: XXXXX (dec., in

1

/

10

°C or. °F)

last digit is the decimal place

(88880

=

Temp.-Overflow)

Reading
temperature
value repeated:

AAmsXXX

XXX = 000 …999 (XXX = number of measuring values)

Reading basic
temperature
range:

AAmb

Output: XXXXYYYY (hex 8-digit, °C or °F)

XXXX = beginning of temp. range

YYYY=

end

of

temp.

range

Reading temp.

sub range:

AAme

same as mb

Changes only via PC software InfraWin

Reading
emissivity:

AAem

Output: XXXX (dez. 0010 … 1000 in ‰)

Emissivity:

AAemXX

XX = (10…99%), 00 = 100% (decimal)

Emissivity:

AAemXXXX

XXXX = (0010 … 1000‰) (decimal)

Exposure time t

90

: AAezX

X = 0 ... 6 (decimal) 0 = intrinsic time constant of the
device
1 = 0.01 s

3 = 0.25 s

5 = 3.00 s

2 = 0.05 s

4 = 1.00 s

6 = 10.00 s

External deletion: AAlx

Simulation of an external deletion contact

Setting of
temperature
sub range:

AAm1XXXXYYYY XXXX (hex 4-digit) beginning of temp. range (°C)

YYYY (hex 4-digit) end of temp. range (°C)