IS 12 / IGA 12 Manual

Instrument Settings 22

3) ENT:

If a parameter is changed with the arrow keys, the new value
must be confirmed by pushing the ENT key. If it is not confirmed
with ENT, the instrument will operate with the previous

parameter value. If no key is pressed for approx. 30 s, the display
changes to the temperature indication.

Limit contacts:

Two limit switches can be set. If the measuring temperature

exceeds the adjusted limit contact temperature, the LED S1 or S2
displays the switch status.

Scale °C or °F:

The LED indicates the temperature scale in °C or °F.

Parameter indicator:

LED’s indicate which pyrometer parameter is selected for reading

or changing.

4.3 Factory settings

Emissivity ( ) = 1,000

Exposure time (t

90

/ s) = 0,00

Clear time (t

CL

) = 0,00

Sub range (from / to) same as temperature range

Analog output (0 / 4 mA) = 0 ... 20 mA

Limit contacts = end of temperature range

Hysteresis = ±2 °C

scanning angle (only IS 12-S and IGA 12-S): 4°
scanning frequency (only IS 12-S and IGA 12-S): 5 Hz

Temperature display (°C / °F) = °C

Interface (RS) = RS232
Address (ADR) = 00
Baud rate (kBaud) = 19.2 kBd
Test current output (10 mA) = off

4.4 Built-in Scanner (IS 12-S and IGA 12-S)

The pyrometers IS 12-S and IGA 12-S with fixed optics are

equipped with a scanning mechanism built into the
pyrometer housing which moves the measuring beam up
and down. In combination with the pyrometer's
maximum value storage (peak picker), the scanner is
optimally used for scanning of thin oscillating wires, for
finding scale-free spots on heavily scaled surfaces or for

measuring small, hot objects whose position is not exactly
determined.
All instruments are equipped with a thru-lens view finder
and an additional laser targeting light for exact
alignment to the position of the measuring object. The thru-lens view finder doesn’t follow the

scanning mirror movement; it always shows the center of the scanning amplitude. The laser
targeting light follows the scanning mirror movement and always shows the position of the
measuring spot. The moving measuring beam does not increase the spot sizes due to the very
fast exposure time of the pyrometers.
The scanning length increases with increasing measuring distance. An overview of the scanning
length at the different distances of the optics is shown in the table.

distance a

scanning distance at

4° scanning angle

a = 80 mm

5.6 mm

a = 160 mm

11.2 mm

a = 250 mm

17.5 mm

a = 660 mm

46 mm

a = 1300 mm

91 mm

a = 5600 mm

391 mm

Scanning
amplitude