IS 12 / IGA 12 Manual

Controls and Instalaltion 15

3 Controls and Installation

3.1 Electrical Installation

The pyrometers are powered by a voltage of either 24 V DC (15 to 40 V DC) or 24 V AC
(12 to 30 V AC), 48 to 62 Hz. Once connected to power, the instrument operations immediately

and needs no warm-up time. To switch off the instrument, unplug the connector.
To meet the electromagnetic requirements, a shielded connecting cable must be used. The
shield of the connecting cable has to be connected only on the pyrometer side to avoid ground
loops.
LumaSense offers connecting cables, but they are not part of

standard scope of delivery. The main connecting cable has wires
for power supply, interface, analog output, external laser switch
and external clear of maximum value storage via contact (see
Chapter 10, Reference numbers) and 12 pin angle connector.

The cable includes a short RS232 adapter cable with a 9 pin
SUB-D connector for direct PC communication. This adapter is
not used in combination with RS485 interface.
For use of the limit contacts, the separate additional cable has

to be used.

connector

connector

for

main for

additional

cable

connection cable for limit switches

3.1.1 Pin assignment for the connector on the back of the pyrometer

For the main connection cable

Pin Color Indication

K

A

B

C

D

J

G

F

E

H

M

L

Pin assignment

of left connector

(side of

male inserts)

K

white

+ 24 V power supply (or 24 V AC) (12 ... 30 V)

A

brown

0

V power supply

L green

+ I

outp.

analog output

B yellow

– I

outp.

analog output

H

gray

external

switch

for

targeting

light

(bridge

to

K)

J pink

external clearing of maximum value storage (bridge to
K)

*

)

or output for switch contact (see Section 5.13)

G

red

DGND

(Ground

for

interface)

F

black

RxD

(RS232)

or

B1

(RS485)

C

violet

TxD

(RS232)

or

A1

(RS485)

D gray/pink B2 (RS485) (bridge to F)

E

red/blue A2

(RS485)

(bridge

to

C)

M orange

Screen only for cable extension
don’t connect at the switchboard

*

)

for setting of clear time to "extern" (see 5.3, clear time for maximum value storage)

For additional limit contacts cable

The instrument is equipped with two independent relay limit switches. These are two separate
switch-over relay contacts, changing its state if the adjusted temperatures are exceeded.