12 wait time, 13 switch contact – LumaSense Technologies IS 12-S User Manual
Page 28
IS 12 / IGA 12 Manual
Settings / parameter descriptions 28
external indicator. If both displays show different readings the external indicator has a wrong
temperature range or input current setting. The test function is switched off automatically after
1 minute and the instrument is working in the measuring mode.
5.12 Wait Time
Using a pyrometer with RS485 it is possible that the connection is not fast
enough to receive the pyrometer’s answer to an instruction of the master. In
this case, a wait time can be set to slow down the data transfer (e.g.: tw = 02
at a baud rate 9600 means a wait time of
2
/
9600
sec).
Note: Only available via interface commands, see Chapter 9, Data format UPP.
5.13 Switch Contact
The pyrometer is equipped with a built-in switch contact which allows you to use the pyrometer
as a thermo switch. This function enables the detection of a hot object in the measuring beam
of the pyrometer. The contact is activated only in combination with a clear time settings “auto“
or clear times 1 s (see Section 5.3 t
CL
/ s (clear times of the maximum value storage)). If
the temperature exceeds 2 °C min. or 1% of the span of the temperature sub range above the
minimum temperature, the supply voltage (pin K) is connected to pin J.
Settings:
00 Bit
...
99 Bit