IS 12 / IGA 12 Manual

Settings / parameter descriptions 28

external indicator. If both displays show different readings the external indicator has a wrong
temperature range or input current setting. The test function is switched off automatically after
1 minute and the instrument is working in the measuring mode.

5.12 Wait Time

Using a pyrometer with RS485 it is possible that the connection is not fast

enough to receive the pyrometer’s answer to an instruction of the master. In
this case, a wait time can be set to slow down the data transfer (e.g.: tw = 02
at a baud rate 9600 means a wait time of

2

/

9600

sec).

Note: Only available via interface commands, see Chapter 9, Data format UPP.

5.13 Switch Contact

The pyrometer is equipped with a built-in switch contact which allows you to use the pyrometer

as a thermo switch. This function enables the detection of a hot object in the measuring beam
of the pyrometer. The contact is activated only in combination with a clear time settings “auto“
or clear times 1 s (see Section 5.3 t

CL

/ s (clear times of the maximum value storage)). If

the temperature exceeds 2 °C min. or 1% of the span of the temperature sub range above the
minimum temperature, the supply voltage (pin K) is connected to pin J.

Settings:

00 Bit

...

99 Bit