IS 310 / IGA 310 Manual

Troubleshooting 17

6 Troubleshooting

Before sending the pyrometer for repair, try to find the error and to solve the problem with the

help of the following list.

Temperature indication too low

Incorrect alignment of the pyrometer to the object

New correct alignment to achieve the max. temperature signal

Measuring object is smaller than spot size

Check measuring distance, smallest spot size is at nominal measuring distance (see 3.3)

Emissivity set too high

Set lower correct emissivity corresponding to the material (see 4.1)

Lens contaminated

Clean lens carefully (see 5.2)

Temperature indication too high

Emissivity set too low

Set higher correct emissivity corresponding to the material (see 4.1)

The measurement is influenced by reflections of hotter machine parts

Try to avoid the influence of the interfering radiation or change measuring position

Measuring errors

Indicated temperature is decreasing during the use of the pyrometer, contamination of

the lens

Clean lens Recommendation: use of air purge

Indicated temperature is decreasing during the use of the pyrometer, although the air

purge unit is used. Probably compressed air is not clean or air failed

Clean the lens and use clean, dry and oil free compressed air

Strong HF-interferences

Correct the connection of the cable shield (see 3.1)

Instrument overheated

Use cooling jacket with air or water cooling