IS 310 / IGA 310 Manual
Controls and Installation 11
3 Controls and Installation
3.1 Electrical Installation
The instruments are supplied by 24 V DC ( 25%) and a ripple < 50 mV. Ensure correct polarity
when connecting the instrument to the power supply. The power consumption (in this case,
4 ... 20 mA) is also the measuring signal. The instrument doesn’t need any time for starting or
preheating and is immediately ready for operation. To switch off the instrument, interrupt the
instrument’s power supply.
Cable colors: white: +24 V DC
green: LED +5 ... 30 V DC
brown: 0 V
yellow: LED 0 V
black: screen
To meet the electromagnetic requirements, a shielded connecting cable must be used. The
shield of the connecting cable is usually only connected on the pyrometer side. If the connecting
cable is extended, the extension cable also needs to be shielded. Do not connect the shield in
the control cabinet to avoid ground loops.
3.1.1 Connection of additional devices
Example for wiring using a digital display with integrated power supply:
Controller
°C
Recorder
Digital display
with built-in
2-wire supply
230 V
~
white
brown
Shield (black)
4 ... 20 mA
+ green
- yellow
Power supply
for LED targeting light
White:
+ 24 V
Brown:
0 V
Black:
Shield
Green:
LED +
Yellow:
LED
-
Example for wiring using an external power supply:
Controller
°C
Recorder
Digital display
230 V
~
4 ... 20 mA
white
brown
Shield (black)
24 V DC
Power supply
+ green
- yellow
Power supply
for LED targeting light
White:
+ 24 V
Brown:
0 V
Black:
Shield
Green:
LED +
Yellow:
LED
-
Note: Additional analyzing instruments, e.g. controllers, recorders, etc. can be
connected in series as shown in drawing above.