2 led targeting light, 3 optics – LumaSense Technologies IS 310 User Manual
Page 12
IS 310 / IGA 310 Manual
Controls and Installation 12
3.1.2 Calculation of the measuring temperature from current output
IS 310
MB 18
Temp. [°C] =
71.875 current [mA] + 362.5
IS 310
MB 23
Temp. [°C] =
93.750 current [mA] + 425
IS 310
MB 25
Temp. [°C] =
87.500 current [mA] + 750
IGA 310
MB 13L
Temp. [°C] =
62.500 current [mA] + 50
IGA 310
MB 15
Temp. [°C] =
62.500 current [mA] + 250
3.2 LED targeting light
The pyrometers are equipped with an LED targeting light for aiming onto the measuring object.
The center of the light marks the center of the spot. For use of this targeting light, a separate
power supply is necessary with 5 ... 30 V DC and a current of I 30 mA.
Measuring is possible while the LED targeting light is on. We recommend using a low supply
voltage to keep the thermal loading of the pyrometer small.
3.3 Optics
The pyrometers are equipped ex works with one of the following optics. These optics focus to a
certain distance, i.e. in these distances each optics achieves its smallest spot size in relation to
the measuring distance. The spot size will change in any other distance (shorter or longer).
Please note that the measuring object must be as least as big as the spot size.
The following drawing and table shows the size of the spots in mm in relation to the measuring
distance. Values between the mentioned data can be calculated by interpolation. The spot size
for measuring distance 0 is the aperture diameter of the optics.
Type
D
a
M
a
1
M
1
a
2
M
2
IS 310
(MB 18)
5.2
250
1.8
500
8.8
1000
23
(MB 23+25)
1
500
7.2
1000
20
(MB 18)
600
4
1000
10.1
2000
26
(MB 23+25)
2
1000
6.8
2000
20
(MB 18)
1400
9
2000
15.1
3000
25
(MB 23+25)
4.5
2000
8.7
3000
16
IGA 310
9
250
2
500
13
1000
35
600
4.5
1000
13.5
2000
36
1400
9
2000
16.8
3000
30
Spot size
Measuring distance
D
a
a
a
2
M
2
M
1
M