beautypg.com

2 introduction, 1 appropriate use, 2 scope of delivery – LumaSense Technologies IS 320 User Manual

Page 9: 3 technical data

background image

IS 320 • IGA 320 Manual

General 9

2 Introduction

2.1 Appropriate use

The IS 320 and IGA 320 digital IMPAC pyrometers are designed for non-contact temperature
measurement of metals, ceramics, or graphite.

2.2 Scope of Delivery

Pyrometer, PC adjustment and evaluation software InfraWin, inspection sheet, and manual.

2.3 Technical Data

Temperature

Ranges:

IS 320:

550 ... 1400 °C (MB 14)

600 ... 1600 °C (MB 16)
650 ... 1800 °C (MB 18)

IGA 320:

300 ... 1300 °C (MB 13)

350 ... 1650 °C (MB 16.5)
400 ... 1800 °C (MB 18)

Sub Range:

Any range adjustable within the temperature range, minimum span 51 °C

Spectral Range:

IS 320:

0.8 ... 1.1 µm

IGA 320:

1.45 ... 1.8 µm

IR Detector:

IS 320:

Silicon photodiode (Si)

IGA 320:

Indium Gallium Arsenide photodiode (InGaAs)

Power Supply:

24 V DC (10 to 30 V DC), ripple must be less than 0.5 V

Power

Consumption:

Max. 1 W

Analog Output:

0 to 20 mA or 4 to 20 mA (linear), switchable

Load:

0 to 500

Switch Contact:

Opto relays; max. 50 V DC, 0.2 A; max. P

max

= 300 mW

Hysteresis:

2 to 20 °C

Digital Interface: RS485 addressable (half duplex), baud rate 1200 up to 38400 Bd

Resolution:

0.1 °C on interface;
< 0.025% of the adjusted temperature sub range at the analog output

Isolation:

Power supply, analog output, and digital interface are galvanically
isolated from each other

Parameters:

Adjustable via interface:
Emissivity , transmittance , exposure time t

90

, clear times for max. / min.

value storage or automatically deletion of max. / min. value storage,
analog output 0 - 20 or 4 - 20 mA (switchable), sub temperature range,

ambient temperature compensation, address, switch contact, hysteresis,
baud rate, wait time t

W

Emissivity :

10.0 to 100.0% adjustable via interface in steps of 0.1%

Transmittance :

10.0 to 100.0% adjustable via interface in steps of 0.1%

Exposure Time

t

90

:

2 ms (with dynamical adaptation at low signal levels);

adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s

Maximum /

Minimum Value
Storage:

Built-in single or double storage. Clearing with adjusted time t

clear

(off; 0.01 s;

0.05 s; 0.25 s; 1 s; 5 s; 25 s), via interface or automatically with the next

measuring object

Uncertainty:

(with =1, t

90

=1 s,

T

amb.

=23°C)

Up to 1500 °C:

0.3% of reading in °C + 1 °C

Above 1500 °C: 0.5% of reading in °C

(Note: the pyrometer must be in operation at least 30 min before these values are valid)

This manual is related to the following products: