Accuracy
Pressure
0.3 psi
±0.1% F.S.
1.0 psi, 5 psi
±0.05% F.S.
15 psi through 3000 psi
±0.025% F.S.
5000 psi, 10,000 psi
±0.035% F.S.
mA
±0.015% of rdg ±0.002mA
Volts
±0.015% of reading ±0.002V
RTD (ohms)
±0.015% of rdg ±0.02 ohms; or ±0.1°C @ 0°C for Pt100
Temperature Effect
No effect on accuracy on all functions from 15°C to 35°C
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C
Optional LPT100A Probe
Meets PT-100 ALPHA 385 Class "A" Specifications
14. Warranty
CONDEC warrants all products against material defects and
workmanship for a period of twelve (12) months after the date of
shipment. Problems or defects that arise from misuse or abuse of the
instrument are not covered. If any product is to be returned, a
"Return Material Authorization" form can be obtained from our
website www.4condec.com under CONTACT CONDEC. You can also
call 1-888-295-8475 to have a form faxed. CONDEC will not be
responsible for damage as a result of poor return packaging. Out of
warranty repairs and recalibration will be subject to specific charges.
Under no circumstances will CONDEC be liable for any device or
circumstance beyond the value of the product.
15. Maintenance
15.1 Replacing Batteries
Replace batteries as soon as the battery indicator turns on to avoid
false measurements. If the batteries discharge too deeply the
CONDEC 3030 will automatically shut down to avoid battery leakage.
Note: Use only AA size alkaline batteries or optional rechargeable
battery pack.
15.2 Cleaning the Unit
Warning
To avoid personal injury or damage to the calibrator, use only the
specified replacement parts and do not allow water into the case.