Warranty, Maintenance, Warning – Condec 3030 User Manual

Page 42: 1 replacing batteries, 2 cleaning the unit

background image

Accuracy

Pressure

0.3 psi

±0.1% F.S.

1.0 psi, 5 psi

±0.05% F.S.

15 psi through 3000 psi

±0.025% F.S.

5000 psi, 10,000 psi

±0.035% F.S.

mA

±0.015% of rdg ±0.002mA

Volts

±0.015% of reading ±0.002V

RTD (ohms)

±0.015% of rdg ±0.02 ohms; or ±0.1°C @ 0°C for Pt100

Temperature Effect

No effect on accuracy on all functions from 15°C to 35°C
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C

Optional LPT100A Probe

Meets PT-100 ALPHA 385 Class "A" Specifications

14. Warranty

CONDEC warrants all products against material defects and
workmanship for a period of twelve (12) months after the date of
shipment. Problems or defects that arise from misuse or abuse of the
instrument are not covered. If any product is to be returned, a
"Return Material Authorization" form can be obtained from our
website www.4condec.com under CONTACT CONDEC. You can also
call 1-888-295-8475 to have a form faxed. CONDEC will not be
responsible for damage as a result of poor return packaging. Out of
warranty repairs and recalibration will be subject to specific charges.
Under no circumstances will CONDEC be liable for any device or
circumstance beyond the value of the product.

15. Maintenance

15.1 Replacing Batteries

Replace batteries as soon as the battery indicator turns on to avoid
false measurements. If the batteries discharge too deeply the
CONDEC 3030 will automatically shut down to avoid battery leakage.

Note: Use only AA size alkaline batteries or optional rechargeable
battery pack.

15.2 Cleaning the Unit

Warning

To avoid personal injury or damage to the calibrator, use only the
specified replacement parts and do not allow water into the case.