Condec 3030 User Manual
Page 40

ST_CLOSE?
Returns the pressure that the switch closed at in the current
pressure units.
ST_OPEN?
Returns the pressure that the switch opened at in the current
pressure units.
ST_DEAD?
Returns deadband of the switch in the current pressure units.
TEMP_UNIT
This command is used to the temperature unit used when
measuring temperature.
The first argument indicates which display to apply the change
to. The second argument is the unit, either CEL for Celsius or
FAR for fahrenheit.
For example:
To set the temperature unit to fahrenheit on the lower
display send TEMP_UNIT LOWER, FAR.
TEMP_UNIT?
Returns the temperature unit, (CEL or FAR) used when
measuring RTDs for each of the 3 displays.
VAL?
Returns the value of any measurement taking place on the upper
and lower display. For example, if the upper display is measuring
5mA, and the lower display is measuring 10V, then VAL? will
return:
5.000000E-03, A, 1.000000E+01, V
ZERO_MEAS
Zeroes the attached pressure module. Enter the zeroing value in
PSI after the command when zeroing an absolute pressure
module.
ZERO_MEAS?
Returns the zero offset or the reference value for absolute
pressure modules.
38