Condec 3030 User Manual
Page 38

FUNC?
Returns the current mode of all displays. For example if the
calibrator is set to [P2] ST on the upper display, [P1] on the
middle, and RTD on the lower, FUNC? Would return:
ST_P2,P1,RTD
HI_ERR
Sets the 100% point for the percent error mode calculation in the
current engineering units.
For example:
To set the 100% point to 100 psi send HI_ERR 100.
HI_ERR?
Returns the 100% point for the percent error mode calculation.
For example:
If the 100% point is set to 100 psi, HI_ERR? would return
1.000000E+02, PSI .
LOCAL
Restores the calibrator to local operation if it was in remote
mode. Also clears LOCKOUT if the calibrator was in lockout
mode.
LOCKOUT
Sending this command sets the lockout state, when the unit is in
REMOTE or goes to remote it prohibits use of the keypad
completely. The lockout state can only be cleared by sending
the LOCAL command.
LO_ERR
Sets the 0% point for the percent error mode calculation in the
current engineering units.
For example:
To set the 0% point to 20 psi send LO_ERR 20.
LO_ERR?
Returns the 0% point for the percent error mode calculation.
For example:
If the 0% point is set to 20 psi, LO_ERR? would return
2.000000E+01, PSI .
36