Condec 3030 User Manual
Page 37
![background image](/manuals/361797/37/background.png)
ERROR _LOOP?
Returns the current state of loop power in percent error mode.
For example:
If you send ERROR_LOOP? It will return ON if loop power
is on in error mode.
ERROR_ MODE
Turns percent error mode on and off.
For example:
To turn on percent error mode send ERROR_MODE ON.
ERROR _ MODE?
Returns the current state of percent error mode.
For example:
If you send ERROR_MODE? It will return ON if the
calibrator is in percent error mode.
ERROR_ PORT
Sets the pressure port for percent error.
For example:
To set the pressure port for percent error to [P1] send
ERROR_ PORT P1.
ERROR _ PORT?
Returns the current pressure port for percent error mode.
For example:
If you send ERROR _PORT?, it will return P1 if the
pressure port in percent error is [P1].
FUNC
Sets the display indicated in argument one to the function
indicated in argument 2.
For example:
To set the lower display to RTD mode send FUNC
LOWER,RTD.
35