3 error messages and corrective actions, Error, Message – Yokogawa DAQWORX User Manual
Page 24: Executable function messages, 3 error messages and corrective actions -5, Error -5, Message -5, Executable function messages -5, 3 error messages and correctve actons, Index

3-5
IM WX1-02E
Deta
led Descr
pt
on of Funct
ons
1
2
3
Index
3.3
Error Messages and Correctve Actons
Error
No.
Message
Correctve Actons
E211
Cannot write to file.
Check if the disk capacity is sufficient or if the file systems is
normal.
E212
Cannot read file.
Check if the file exists and is supported by the software or if the
file system is normal.
E213
Cannot open file.
Check if the file exists and is supported by the software or if the
file system is normal
E401
Communication error.
Check if the recorder connected for communication is powered on
and if the cable is properly connected. Also check the following
items according the the communication type.
• For Ethernet
Check if address settings are correct; the TCP/IP protocol is
installed in Windows; the Ethernet card is properly installed.
• For RS-232 and RS-422-A
Check if the baud rate settings match; the port (COM1 to COM9)
settings match, the address settings are correct (RS-422-A); the
serial port of the PC is active and the appropriate cable is being
used.
E402
Communication timeout.
-
E403
Cannot open a communication port.
Same as E401.
E501
Invalid license number.
Please reinstall the software.
E1001
Login error.
Check the user name, user ID, and password on the DX-P.
E1002
Invalid password.
-
E1003
Password is not valid.
-
E1004
Password expired.
-
Please set up a password again on the DX-P.
E1010
Execution of aprocess failed.
Check whether an executable function exists, or whether its
files are damaged. If this error appears frequently, reinstall the
software.
E1011
Execuition of a service failed.
Check whether an executable function exists, or whether its
files are damaged. If this error appears frequently, reinstall the
software.
E1012
Login prohibited because another user is
logged in.
-
Message
No.
Message
M1201
Model determination was successful.
M1210
Setting changes saved before execution.
Executable Functon Messages
No.
Message
Correctve Actons
W[631]
Data Lack
Reduce the number of acquired data points or connected
instruments, or lengthen the scan interval.
E[673]
Cannot open communication
Same as E401.
E[674]
Communication error
Same as E401.
E[675]
Communication time out
Same as E401.
E[700]
Cannot login
Same as E1013.
E[701]
Cannot Get DXP inoformation
Check whether the communication status and connected
instruments matches those specified in the software. If they do
not, perform model determination again.
I[606]
Recovery Communication
Connection recovered.