Cannot take measurements (when ACTIVE LED does not blink).

Possible Problem

Corrective Action

Refer to

Page

Cable not connected.

Check CAN cable connections.

Wiring incorrect.

Check pin assignments and wiring methods.

Using Low Speed CAN.

LS CAN not supported.

Terminator settings not correct.

Enter terminator settings as necessary (dip
switch no. 8).

Bit rate not appropriate.

Change to an appropriate bit rate setting.

P. 13

All message numbers are OFF.

Change message format and ID to an
appropriate setting.

P. 15

Message ID incorrect.

Change message format and ID to an
appropriate setting.

P. 15

Malfunction.

Servicing required.

Cannot take measurements (when ACTIVE LED blinks).

Possible Problem

Corrective Action

Refer to

Page

Channel’s message numbers are OFF.

Select a message number.

P. 16

Data type incorrect.

Change the data type to an appropriate setting.

P. 16

Start bit position not correct.

Change Start bit position to an appropriate
setting.

P. 17

Bit length not correct.

Change bit length to an appropriate setting.

P. 17

Endian not correct.

Change Endian to an appropriate setting.

P. 17

Measured value is non-numerical when
data type is Float or Double.

If a non-numerical value is received, Invalid
is displayed.

Malfunction.

Servicing required.