beautypg.com

Yokogawa Standalone MW100 User Manual

Page 11

background image

11/17

When Setting Difference between Channels Computation

Syntax SR

p1,p2,p3,p4,p5,p6,p7

p1

Channel range (001–060)

p2

Input mode (COM)

p3

Measurement range (CAN)

p4

Lower limit of span (-30000–30000)

p5

Upper limit of span (-30000–30000)

p6

Computation mode (DELTA)

p7

Reference channel number

(001–060)

Query SR[

p1]?

Example)

Set the MATH type for channel 10 to difference computation between

channels with channel 1 (the reference channel), set the input type to COM(CAN),
the lower limit of span to -5000, and the upper limit of span to 5000.

SR 010,COM,CAN,-5000,5000,DELTA,001
Explanation) Specify

measurement channels on p1.


When Setting Scaling

Syntax SR

p1,p2,p3,p4,p5,p6,p7,p8,p9,p10

p1

Channel range (001–060)

p2

Input mode (COM)

p3

Measurement range (CAN)

p4

Lower limit of span (-30000–30000)

p5

Upper limit of span (-30000–30000)

p6

Computation mode (SCALE)

p7

Scaling lower limit (-30000–30000)

p8

Scaling upper limit (-30000–30000)

p9

decimal place (0–4)

p10

Unit (max 6 characters)

Query SR[

p1]?

Example)

Convert the CAN signal measured on channel 2 to DC current. Set the

lower limit of span to 1000, upper limit of span to 5000, lower limit of scaling to 0.00 A,
and upper limit of scaling to 5.00 A.

SR 002,COM,CAN,1000,5000,SCALE,0, 500,2,A
Explanation)

Specify measurement channels on p1.

For the characters that can be used for p10, see the user specified characters in
section 1.3 of the MW100 Communication Command Manual (IM MW100-17E).