2 GENERAL SPECIFICATIONS

The ISC40 inductive conductivity sensor is suitable for use with the Yokogawa inductive

conductivity analyzers.

2.1 Measuring elements

: Toroïds with high permeability magnetic material

Pt1000 or 30k temperature element

2.2 Materials

Wetted parts sensor

Body ISC40*-G*

: 30% glass filled PEEK, FDA approved

Body ISC40*-T*

: PFA, FDA approved

Non-wetted parts sensor

Sealing gasket

: Viton

Thread part

: AISI 316 SS

Options for sensor

All options except /TFD : AISI 316 SS and O-ring material as wetted part

/TFD

: AISI 316 SS as non-wetted part

TFM and Kalrez as wetted part

2.3 Functional specifications (at 25°C)

Temperature element : Pt1000 to IEC 751

30k NTC

Installation factor

: 1.88 cm-1 nominal for PEEK sensor

3.00 cm-1 nominal for PFA sensor

Actual installation can change this factor.

If there is less than 25mm spacing between sensor and holder,

in-situ calibration is necessary to meet the specified accuracies

(see fig. 1)

Fig 1: Actual installation factor as function of spacing around the sensor

Note: The ISC40 temperature sensor is designed for cell compensation and for indication.

It is NOT designed for process temperature control.

2.4 Dynamic specifications

Response time

: t

90

< 5 min. for PEEK sensor

conductivity

:

t

90

< 10 min. for PFA sensor

D

non conductive

piping

conductive

piping

D i n m i l l i m e t e r s

0 10 20 30 40 50

0.90

Correction factor (x nominal C.C.)

0.95

1.30

1.25

1.20

1.00

1,05

1.10

1.15

O n l y f o r P E E K