Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual
Page 89

IM 12D06A03-01E
Appendix 2-17
4. Temperature Compensation Coefficient Setting Example
5. Calibration in Air example (Air Set)
*HOLD
*CODE 04
*AIR
*START
*END
Confirm that display is 0.0 μS/cm
*TEMP
*SERV
¼
YES
NO
NO
YES
YES
YES
ENT
example5E.eps
*OUTP
*HOLD
*NaCl
*TC
*SERV
TC 1.24
*TEMP
CODE 21
α
=
K
t -
K
ref
T - T
ref
K
ref
100
x
T = Measured temperature ( °C )
Kt = Conductivity coefficient at temperature T ( °C )
Tref = Reference temperature ( 25 °C )
Kref = Conductivity coefficient at temperature Tref
The temperature compensation coefficient is calculated as follows:
MODE
Escape
ENT
YES
NO
NO
NO
YES
YES
ENT
¼
example4E.eps
See also other documents in the category Yokogawa Sensors:
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA210A (70 pages)
- EJA430A (78 pages)
- EJA130A (4 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJX430A (76 pages)
- EJA430E (96 pages)
- EJA430E (41 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)