Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual
Page 86

IM 12D06A03-01E
2-14 Appendix
Sample 4
Concentration of hydrochloric acid
Meas. range
30-40%
Reference temp.
35°C
Temp. compensation
Enter temperature compensation coefficient,
Relate concentration to conductivity by table
START
Service code 20
Service code 31
Service code 35
Select 1
0% output --> 30 %
100% output --> 40%
Service code 55
TEMP
Calibration in Air
Calibration
35 °C (Default)
NaCl --> NO
TC --> YES ( here use [MODE] key to escape )
Calibrate conductivity value in actual operation
e.g. 34% HCl corresponds to conductivity 772 mS/cm
Select 1
Enter conductivity values corresponding to output current values.
Table (below) relates output current to conductivity at 35°C
Service code 21
TC=1.24 (data from customer)
Output
current
(%)
0
20
40
60
80
100
HCl conc.
(%)
30
32
34
36
38
40
Conductivity
(mS/cm)
845
840
772
731
688
644
sample4E.eps
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJX910A (9 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (96 pages)
- EJA430E (41 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)