beautypg.com

Calibration, 1 when is calibration necessary, Calibration -1 – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 63: 1 when is calibration necessary? -1

background image

IM 12D06A03-01E

Calibration 6-1

6. CALIBRATION

6-1 When is calibration necessary?

Calibration of conductivity instruments is normally not required, since Yokogawa delivers a wide range

of sensors, which are factory calibrated traceable to NIST standards. The cell constant values are nor-

mally indicated on the top of the sensor or on the integral cable. These values directly can be entered in

service code 03 (section 5-3-1). If the cell has been subjected to abrasion (erosion or coating) calibration

may be necessary. In the next section two examples are given. Alternatively calibration may be carried

out with a simulator to check the electronics only.

NOTE:

During calibration the temperature compensation is still active. This means that the readings are

refered to the reference temperature as chosen in service code 20 (section 5-3-4, default 25 °C).

alibration is normally carried out by measuring a solution with a known conductivity value at

a known temperature. The measured value is adjusted in the calibration mode. On the next

pages the handling sequence for this action is visualised. Calibration solutions can be made up

in a laboratory. An amount of salt is dissolved in water to give a precise concentration with the

temperature stabilised to the adjusted reference temperature of the instrument (default 25 °C). The

conductivity of the solution is taken from literature tables or the table on this page.

Alternatively the instrument may be calibrated in an unspecified solution against a standard instrument.

Care should be taken to make a measurement at the reference temperature since differences in the type

of temperature compensation of the instrument may cause an error.

NOTE:

The standard instrument used as a reference must be accurate and based on an identical

temperature compensation algorithm. Therefore the Model SC72 Personal Conductivity Meter of

Yokogawa is recommended.

Typical calibration solutions.

The table shows some typical conductivity values for sodium-chloride (NaCl) and Potassium cholide

(KCl) solutions which can be made up in a laboratory.

Table 6-1. NaCl values at 25 °C (IEC 60746-3)

Table 6-2. KCl values at 25 °C

Weigth %

molal
(m)

mg of KCl /
kg of solution

Conductivity

0.3

0.001

74.66

0.1469 mS/cm

0.5

0.002

149.32

0.2916 mS/cm

1

0.005

373.29

0.7182 mS/cm

3

0.01

745.263

1.4083 mS/cm

5

0.1

7419.13

12.852 mS/cm

10

1.0

71135.2

111.31 mS/cm

The table is derived from the Standards laid down in ‘International
Recommendation No. 56 of the Organisation Internationale de
Métrologie Legale’.

Weigth %

mg/kg

Conductivity

0.001

10

21.4 μS/cm

0.003

30

64.0 μS/cm

0.005

50

106 μS/cm

0.01

100

210 μS/cm

0.03

300

617 μS/cm

0.05

500

1.03 mS/cm

0.1

1000

1.99 mS/cm

0.3

3000

5.69 mS/cm

0.5

5000

9.48 mS/cm

1

10000

17.6 mS/cm

3

30000

48.6 mS/cm

5

50000

81.0 mS/cm

10

100000

140 mS/cm