Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual
Page 84

IM 12D06A03-01E
2-12 Appendix
Sample 2
NaOH concentration
Meas. range
0-10%
Reference temp.
25°C Temp. compensation
Use existing (temperature, concentration) matrix
Service code 55
TEMP
Calibration in Air
Calibration
Service code 22
START
Output Current (%)
0
10
30
60
100
NaOH conc (%)
0
1
3
6
10
Conductivity (mS/cm)
0
53
145
256
359
Enter conductivity values corresponding to output current values.
Existing table (below) relates output current to conductivity at 25°C
Service Code 20
25°C (default)
Service Code 31
Service Code 35
Select 1
0% output --> 0 %
100% output --> 10%
NaCl --> NO
TC --> YES ( here use [MODE] key to escape )
Matrix code 8
Calibrate conductivity value in actual operation
e.g. 3% NaOH corresponds to conductivity 145 mS/cm
(see above table)
sample2E.eps
Select 1
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (52 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJX910A (83 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (41 pages)
- EJA430E (96 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)