Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual
Page 59

IM 12D06A03-01E
Parameter setting 5-17
Code Display
Function
Function
detail
X
Y
Z
Default
values
mA Outputs
30
Not
used
31
*OUTP.F
mA output functions
Linear
0
0 Linear
Table
1
32
*BURN
Burn function
No burnout
0
0 No Burn.
Burnout
downscale
1
Burnout
upscale
2
Pulse
burnout
3
33, 34
Not used
35
*TABLE
Output table for mA
*0%
Linearisation table for mA in 5% steps.
*5%
The measured value is set in the main
*10%
display using the
keys, for
...
each of the 5% interval steps.
...
Where a value is not known, that value may
*95%
be skipped, and a linear interpolation will
*100%
take
place.
36-49
Not
used
mA
Concentration Output function is done in de following order:
• Set *OUTP.F. (Service Code 31) to table
• Set the Concentration range in % (Service Code 55)
• Set table values (%output and Conductivity values) in *TABLE (Service Code 35)
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA120A (47 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJA210E (52 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJX910A (9 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (96 pages)
- EJA430E (41 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)