APPENDIX 1. LIST OF PARAMETERS FOR EACH BLOCK OF THE EJA
A1.3 Transducer Block Parameter List (Slot=3)
Index
16
Parameter
Block Object
Write
Mode
--
Initial Value
Description
Information on this block such as Block Tag, DD Revision,
Execution Time.
24
Pressure Raw Value
--
Variable
This parameter contains the raw sensor value.
The uncalibrated measurement value from the sensor.
25
Sensor Upper Limit
--
Measurement
span
This parameter contains the sensor upper limit value.
26
Sensor Lower Limit
--
Measurement
span
This parameter contains the sensor lower limit value.
29
Calibration Span Min
--
Minimum span
of capsule
The minimum calibration span value allowed.
34
Measured Value
--
Variable
This parameter contains the measured value and status
available to the AI1 Block.
36
Transmitter Type
O/S
Pressure
This parameter contains the application of the pressure device.
Pressure, Flow, Level, Volume, Special
44
Temperature Unit
O/S
˚C
This parameter contains the temperature units.
˚C/˚F
19
Strategy
AUTO 0
The strategy field can be used to identify a grouping of blocks.
This data is not checked or processed by the block.
20
Alert key
AUTO 0
The identification number of the plant unit. This information
may be used in the host for sorting alarms, etc.
21
Target Mode
AUTO AUTO
Set the Target of block mode (MODE_BLK) to Auto or O/S
according to the Write Mode of the parameter to be set or changed.
30
Pressure Raw Value Unit
O/S
Specified at the
time of order
This parameter contains the engineering unit’s index code for
the calibration values. SENSOR_UNIT must be a subset of the
interchangeable part of the Pressure unit.
Pa, GPa, MPa, kPa, mPa, microPa, hPa, bar, mbar, torr, atm,
psi, psia, psig, g/cm
2
, kg/cm
2
,
inH
2
O, inH
2
O(4degC), inH
2
O(68degF), mmH
2
O,
mmH
2
O(4degC), mmH
2
O(68degF), ftH
2
O, ftH
2
O(4degC),
ftH
2
O(68degC), inHg, inHg(0degC), mmHg, mmHg(0degC)
23
Alarm Sum
--
For future use.
27
Upper Calibration Point
O/S
Max range
The highest calibrated value.
28
Lower Calibration Point
O/S
0
The lowest calibrated value.
18
TAG
AUTO Space
The user description of the intended application of the block.
31
Trimmed Value
--
Variable
This parameter contains the sensor value after the trim
processing.
32
Sensor Type
--
Resonant
The type of sensor. 126: Resonant
43
Sensor Temperature
--
Variable
Temperature value of transducer.
35
Measured Value Unit
O/S
%
This parameter contains the engineering unit’s index code for
the primary value.
All units which are supported in Profile are available.
22
Mode Block
--
The mode parameter is a structured parameter composed of
the actual mode, the normal mode, and the permitted mode.
Actual: Indicates the current operating condition.
Permit: Indicates the operating condition that the block is
allowed to take.
Normal: Indicates the operating condition that the block will
usually take.
17
Static Revision No
--
2
The revision level of the static data associated with the function
block. The revision value will be incremented each time a static
parameter value in the block is changed.
TA0103-01.EPS
45
Pressure Value
--
Variable
This parameter contains the Pressure value before scaling
calibrated range.