15) burst mode, 16) multidrop mode, 16) burst mode -16 – Yokogawa EJA115 User Manual
Page 23
IM 01C22T01-01E
2-16
2. HART COMMUNICATOR OPERATION
F0233.EPS
EJA:
Enter meter value
1.000
HELP
ABORT
9
9
8
Enter the reading of the voltmeter
(1.010), and press ENTER (F4).
(The output of the transmitter
changes.)
DEL
ENTER
(ENTER)
(ENTER)
‘1 . 0 1’
EJA:
Scaled output: 1.000
equal readout
device?
1 Yes
2 No
HELP
ABORT
10
DEL
ENTER
Voltmeter reading: 1.010
Because the reading on the
voltmeter is 1.000, select Yes and
press ENTER (F4).
If the reading is not 1.000, select
No. Repeat steps and until
the voltmeter reads 1.000 V.
Voltmeter reading: 1.000
11
Press OK (F4). The output signal of
100% is output.
EJA:
Setting fld dev
output to 20mA
HELP
SET
ABORT
OK
EJA:
Enter meter value
5.000
HELP
DEL
ABORT
ENTER
12
EJA:
Scaled output: 5.000
equal readout
device?
1 Yes
2 No
HELP
DEL
ABORT
ENTER
13
EJA:
NOTE-Loop may be
returned to automatic
control
HELP
ABORT
14
Enter the reading of the voltmeter
(5.210), and press ENTER (F4).
DEL
OK
(ENTER)
“Returning fid dev to
original output”
(OK)
(ENTER)
‘5 . 2 1’
(OK)
Voltmeter reading: 5.210
Select Yes and press ENTER (F4).
Press OK (F4).
Voltmeter reading: 5.000
(16) Burst Mode
The transmitter continuously sends the data stored in it
when the burst mode is set on. Either one of measured
pressure variable, % output value, or 4 to 20 mA
output value can be selected and sent. The data is sent
intermittently as a digital signal at 75 ms intervals
when the transmitter is set in the burst mode. There-
fore, communication by the HART simultaneous
communicator is also possible.
Setting of Burst Mode
F0234.EPS
Call up the Burst option, and set
the data to be sent.
• Pressure variable (PV)
• % output value (% range/current)
• 4 to 20 mA output value (Process
vars/crnt)
EJA:
Burst option
********
PV
% range/current
Process vars/crnt
HELP
DEL
ESC
ENTER
EJA:
Burst mode
Off
On
Off
DEL
DEL
ESC
ENTER
Call up the Burst mode and set to
On.
Then, Press SEND (F2).
HELP
(ENTER)
(ENTER)
(SEND)
1. Device setup
4. Detailed setup
3. Output condition
4. HART output
3. Burst mode, 4. Burst option
To Release from the Burst Mode:
Call up the
Burst mode
display, and set to
OFF
.