Advisory and semi-automatic modes, Dvisory, Automatic – ZOLL M Series CCT Defibrillator Rev A User Manual
Page 35: Odes
Configuration Option Tables
31
Advisory and Semi-automatic Modes
Feature
Options
Default
Energy Level: Shock 1
Damped Sine Wave (DSW): 1-10, 15, 20, 30,
50, 75, 100, 150, 200, 300, 360 J
Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120,
150, 200 J
Manual DSW: 200 J
Manual Biphasic: 120 J
AED DSW: 200 J
AED Biphasic: 120 J
Energy Level: Shock 2
Damped Sine Wave (DSW): 1-10, 15, 20, 30,
50, 75, 100, 150, 200, 300, 360 J
Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120,
150, 200 J
Manual DSW: 300 J
Manual Biphasic: 150 J
AED DSW: 200 J
AED Biphasic: 120 J
Energy Level: Shock 3
Damped Sine Wave (DSW): 1-10, 15, 20, 30,
50, 75, 100, 150, 200, 300, 360 J
Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120,
150, 200 J
Manual DSW: 360 J
Manual Biphasic: 200 J
AED DSW: 360 J
AED Biphasic: 200 J
Auto Analyze 3 Times
Yes, No, Only in Semi-Auto
Manual: No
AED: Only in Semi-Auto
“NO SHOCK ADVISED” Prompt
“NO SHOCK ADVISED”, “CHECK PULSE”,
BOTH
“NO SHOCK ADVISED”
“CHECK PATIENT” Prompt
“CHECK PATIENT”, “PRESS ANALYZE”
“CHECK PATIENT”
Display “DO CPR” Message
Yes, No, Only in Semi-Auto
Manual: No
AED: Only in Semi-Auto
Display ECG in Semi-Auto Mode
Yes/No
Yes
Store to Card in Auto Monitor Mode
Yes/No
Yes
Auto Charge in Advisory Mode
Yes/No
Yes
Record to Card in Advisory Mode
None, ECG Only, Voice and ECG
None
Enable Voice in Advisory Mode
Yes/No
No
Display Heart Rate in Semi-Auto Mode
Yes/No
No
Auto-Analyze on Power Up
Yes/No
No
Enable 12-Lead in Semi-Auto Mode
Yes/No
Yes
AED Defib Ready Hold Time
15 seconds, 60 seconds
15 seconds