Remove card prompt at pwr off, Sync defib excluded, Ptca settings – ZOLL M Series CCT Defibrillator Rev A User Manual
Page 29: Power up patient type (nibp), Emove, Rompt, Efib, Xcluded, Ptca s, Ettings
Configuration Settings
25
the “Allow Card Erase” option must be set to “Yes”, and the “Record to Card in Advisory
Mode” option must be set to “ECG only” or “Voice and ECG”; otherwise, this option is ignored.
If set to “Yes”, the unit will display the message “Erase Card -- Yes/No?” when the user turns
the Selector Switch to the OFF position and the following criteria are met:
1. A data card is installed in the M Series CTT unit.
2. The unit has NOT performed a defibrillation, other than a self test, since power-up.
Note: The M Series CCT unit can also be configured to display the “Erase Card -- Yes/
No?” prompt if the card contains only synchronized cardioversion data. See the
“Sync Defib Excluded” option for more information.
3. The unit has NOT delivered a pacing current greater than 25 mA since power-up.
If the user selects “Yes” at the prompt within 15 seconds, the card will be erased and the unit
will subsequently shut off. If the user selects “No”, or 15 seconds is exceeded without a
response, the unit will shut off without erasing the card.
Remove Card Prompt at Pwr Off
Sets the M Series CCT unit to prompt the user to remove the data card at power-off if the card
contains any information other than self test, and optionally, synchronized cardioversion data.
If set to “Yes”, the unit will display the message “Remove Card” when the user turns the
Selector Switch to the OFF position and the following criteria are met:
1. A data card is installed in the M Series CCT unit.
2. The unit HAS performed a defibrillation, other than a self test, since power-up.
Note: The M Series CCT unit can be configured NOT to display the “Remove Card”
prompt if the card contains only synchronized cardioversion data. See the “Sync
Defib Excluded” option for more information.
3. The unit HAS delivered a pacing current greater than 25 mA since power-up.
When this option is enabled, the “Remove Card” message will be displayed for 10 seconds,
then the M Series CCT unit will shut off.
Sync Defib Excluded
Allows the exclusion of synchronized cardioversion from the criterion of a defibrillation event,
for the purposes of the “Erase Card Prompt at Pwr Off” and “Remove Card Prompt at Pwr Off”
configuration settings.
When set to “Yes”, if the M Series CCT unit has only discharged self test or synchronized
cardioversion defibrillations and has not delivered pacing current greater than 25 mA, the unit
displays the “Erase Card -- Yes/No” prompt at power-off (if so configured). Otherwise, the unit
displays the “Remove Card” prompt at power-off (if so configured).
PTCA Settings
Allows review of, or changes to Percutaneous Transluminal Coronary Angioplasty (PTCA)
settings if the M Series CCT includes the 12-Lead Reperfusion Algorithm option.
With PTCA Settings highlighted and Set/Review selected, press the Change Value softkey to
set the following parameters:
•
Message Enabled: Determines whether or not the M Series CCT unit prints a PTCA
message (if appropriate).
•
Threshold:
Sets the threshold, in mV, for issuing the PTCA message. Only
available if the Message Enabled option is set to “Yes”.
Power Up Patient Type (NIBP)
Sets the patient type (adult, pediatric, or neonatal) for NIBP measurements when the unit is
powered up.