Daylight savings, Card capacity message enabled, Erase card prompt at pwr off – ZOLL M Series CCT Defibrillator Rev A User Manual
Page 28: Aylight, Avings, Apacity, Essage, Nabled, Rase, Rompt
M S
ERIES
CCT C
ONFIGURATION
G
UIDE
24
Highlight the Clock Synchronization option and press the Change Value softkey to display the
following screen:
Press the Change Value softkey to scroll through the available values for each parameter.
Press the Next Item softkey to move to the next parameter in the table.
Press the Return softkey to save the values entered and return to the main configuration menu.
You can set the values for the following Clock Synchronization options:
•
Time Zone:
Selects the time zone in which the M Series CCT will be used.
Choices include: Azores, Middle Atlantic, S. America Eastern, AST
Atlantic, EST Eastern, CST Central, MST Mountain, PST Pacific,
AKST Alaska, HST Hawaiian, Samoa, Dateline, GMT Greenwich,
CET Central Europe, EET Eastern Europe, MSK Moscow, Arabian,
Afghanistan, West Asia, Central Asia, Southeast Asia, AWST
Australian West, Korean, ACST Australian Cent, AEST Australian
East, Central Pacific, or New Zealand
•
Dial Location:
Sets the National Institute of Science and Technology (NIST) dialing
location to either Hawaii or Colorado. This is the location called by
the M Series CCT to obtain accurate time information for setting its
internal clock.
•
Prefix:
Affixes up to eight user selectable digits or special characters (+, &, !,
*, or comma) to the start of the dialing location phone number.
Daylight Savings
When set to “Yes” and the user dials the selected NIST location, the unit will automatically
adjust the time for Daylight Savings based on the assumption that DST starts on the first
Sunday in April at 1:59:59 AM (setting the time ahead by one hour) and ends on the last Sunday
in October at 1:59:59 AM (setting the time back by one hour).
Card Capacity Message Enabled
When set to “Yes” and data recording is enabled with a data card present, the M Series CCT
will display “CARD LOW” messages when the remaining storage capacity of the card reaches
30, 15, and 5 minutes.
To enable this option, the “Report Memory Card Errors” option must be set to “Yes”; otherwise,
the option is ignored.
Erase Card Prompt at Pwr Off
Sets the M Series CCT unit to allow the user to erase the data card at power-off if the card
contains only self test, and optionally, synchronized cardioversion data. To enable this option,