Store to card in auto monitor mode (aed models), Auto charge in advisory mode, Record to card in advisory mode – ZOLL M Series CCT Defibrillator Rev A User Manual
Page 17: Enable voice in advisory mode, Display heart rate in semi-auto mode, Display lead on power-up, Frequency response, Qrs volume on power-up, Tore, Onitor
Configuration Settings
13
Store to Card in Auto Monitor Mode (AED models)
Sets unit to automatically store patient’s ECG trace and voice data to the PCMCIA card when:
•
ECG
leads are attached to the patient
•
MFE pads are not connected
•
M Seri
es CCT unit is on
If disabled, the M Series CCT unit will not record any data to the PCMCIA card unless MFE
pads are connected to the patient.
Auto Charge in Advisory Mode
When enabled and a shockable rhythm is detected after pressing the ANALYZE button, the
unit will automatically charge the defibrillator to the pre-configured or user selected energy
setting.
Record to Card in Advisory Mode
Sets the unit to record ECG data (ECG only), ECG and voice data (Voice and ECG), or nothing
(None) to the memory card.
Enable Voice In Advisory Mode
Enables or disables voice prompts when operating in Advisory or Manual operating modes. If
enabled, the voice prompts will sound at the appropriate time when in Advisory mode or
Manual mode.
Display Heart Rate in Semi-Auto Mode
Sets the device to display or not display the patient heart rate on the monitor while in
semiautomatic mode.
Display Lead on Power-Up
Allows the user to select any one of the following Leads to be displayed on power-up: Paddles/
Pads, I, II, or III. When the unit powers up in semi-automatic mode, the option is ignored. The
unit will power up in Lead II, and switch leads automatically between leads II and MFE for
Auto Monitor and Auto Defib modes, respectively. When the unit enters Manual Mode (Manual
Mode override sequence is completed), the lead switches to the setting specified in the "Display
Lead on Power Up" option.
Frequency Response
Selects the ECG filter bandwidth to be used during ECG monitoring. The frequency of the
device can be set to either help reduce noise, help with the diagnostic response, or set as a
default response.
Enable 12-Lead in Semi-Auto Mode (AED with 12-Lead)
If set to “Yes,” 12-lead analyses, transmissions and print-outs are available in semi-automatic
(AED) Auto Monitor mode (ECG leads on the patient with MFE Pads disconnected), depending
on other configuration option settings. If set to “No,” 12-lead functions are available in Manual
Mode only.
QRS Volume on Power-Up
Turns the default QRS beeper volume to midrange or off following power up of the unit.
If set to “Off in Semi-Auto Mode,” the QRS beeper volume turns off in semi-auto mode. If the
unit is changed to manual mode, the beeper volume sets to midrange.