Set Lead Groups
This setting allows the user to configure two custom lead groups as described in the 12-Lead
ECG Monitoring insert (part number 9650-0215-01). Pressing the Change Value softkey will
cause the following screen to appear:
Pressing the Prev Lead softkey will select the previous lead.
Pressing the Next Lead softkey will select the next lead.
Pressing the Change Lead softkey will change the selected lead. Repeatedly pressing this key
will cycle the selected lead through all twelve possibilities.
Pressing the Enter softkey will return to main configuration menu.
12 Lead Freq Response
When set to “0.05-150 Hz 4x3” the 12-Lead Filter Setting defaults to 0.05-150 Hz 4x3.
When set to “0.05-150 Hz Cont.” the 12-Lead Filter Setting defaults to 0.05-150 Hz Cont.
When set to “0.05-40 Hz 4x3” the 12-Lead Filter Setting defaults to 0.05-40 Hz 4x3.
Refer to the 12- Lead ECG Monitoring Operator’s Guide insert (part number 9650-0215-01)
for details on each setting.
Lead Group Default
When set to “Standard” the 12-Lead Group Setting defaults to Standard.
When set to “Custom Group 1” the 12-Lead Group Setting defaults to the setting defined under
Custom Group 1.
When set to “Custom Group 2” the 12-Lead Group Setting defaults to the setting defined under
Custom Group 2.
Refer to the 12- Lead ECG Monitoring Operator’s Guide insert (part number 9650-0215-01)
for details on each setting.
Print 12 Lead Interpretation
When set to “Yes” the 12SL interpretative statements will print following the acquisition of a
12-lead ECG. This setting affects the strip chart and summary report printouts as well as the fax
transmission.
Defib Default to Pads
When set to “Yes” switching the unit to DEFIB mode will set the active lead to Pads/Paddles
regardless of the configuration setting for “Display Lead on Power-Up.”