Pace Rate Setting on Power-Up

Sets the default pace rate of the M Series CCT unit on power-up, between 30 ppm and 102 ppm

(increments of 4 ppm).

Display Implanted Pace Spikes

The M Series CCT is capable of detecting and indicating on the display pacemaker signals from

a patient with an implantable pacemaker. Selecting “Yes” causes a pacemaker marker to be

displayed on the ECG trace whenever implanted pacer stimuli are detected. The pacemaker

spike detector is always active in ECG leads, independent of this configuration setting. This

setting controls spike display only.

Gain On Power-Up

Sets the default SIZE of the displayed ECG signal (AGC, 0.5 cm/mV, 1 cm/mV, 1.5 cm/mV,

and 2 cm/mV) when the device is powered-up.

Retain SYNC after Defib

Causes the unit to remain in SYNC mode after a synchronized cardioversion shock. The unit

will remain in SYNC mode until the SYNC button is pressed again or the unit is switched out of

Defib mode.

Auto Generate Strips

Sets the strip chart recorder to automatically begin printing after a defibrillator discharge or

whenever alarms are triggered while in manual mode. (Unit must be equipped with the optional

strip chart recorder feature.)

Print 3 Leads When Leads Sel

When enabled, the strip chart recorder will automatically print three (3) simultaneous leads of

the patient ECG when Leads are selected and a five (5) lead or 12-lead ECG cable is in use.

Report Memory Card Errors

Allows the user to disable memory card error messages and voice prompts.
If this option is set to "No", all error messages (i.e., “INSERT CARD", "REPLACE CARD",

"CARD FULL”, and “CARD LOW < n MINUTES”) related to the use of the memory card will

be suppressed. However, if the unit is in manual mode, or the unit is non-AED, the option is

ignored when the "Record to Card in Advisory Mode" option is set to "None" (no data

recording).

Auto Analyze on Power Up (AED Semi-Automatic Mode)

When set to “Yes”, the unit will automatically begin an ECG rhythm analysis as soon as the

device is turned on and pads are properly connected. An “ANALYZING ECG” message will be

displayed for 5 seconds and a “STAND CLEAR” message is displayed and announced

indicating the start of the analysis. If the Auto Analyze 3 Times option is enabled, the analysis

will be the first of a three analysis sequence.
Note:

The unit must be turned off for 10 seconds or more for the analysis to auto -start when

the device is turned on.

Auto Transmit After 12 Lead

When enabled, the unit will automatically enter the transmission setup screen following

completion of a 12-lead analysis.
To fax using 2x6 format, this must be set to “Yes.”