ZOLL AED Plus Rev V User Manual
Page 43

ZOLL AED PLUS Administrator’s Guide
33
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
<5% Ut (>95% dip
in Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95%dip
in Ut) for 5 sec
Not applicable
Not applicable
Not applicable
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment
NOTE Ut is the a.c. mains voltage prior to application of the test level.
The AED Plus is intended for use in the electromagnetic environment specified below. The
customer or the user of the AED Plus should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the AED
Plus, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bands
a
3 Vrms
d = 1.17
10 Vrms
150 kHz to 80 MHz
in ISM bands
a
10 Vrms
d = 1.20
P
P