ZOLL AED Plus Rev V User Manual
Page 17

ZOLL AED PLUS Administrator’s Guide
7
ADULT PADS.
AED Plus has detected adult electrode pads connected to it and
adjusted defibrillation energy settings to adult levels.
PEDIATRIC PADS.
AED Plus has detected pediatric electrode pads connected to it
and adjusted defibrillation energy settings to pediatric levels.
DON’T TOUCH PATIENT,
ANALYZING.
Do not touch victim, an ECG rhythm analysis is in progress or
about to begin.
SHOCK ADVISED.
ECG rhythm analysis has detected the presence of ventricular
fibrillation or shockable ventricular tachycardia.
NO SHOCK ADVISED.
ECG rhythm analysis has detected a rhythm that is not treatable
by defibrillation.
ANALYSIS HALTED. KEEP
PATIENT STILL.
ECG rhythm analysis has been halted due to excessive ECG
signal artifact. Stop any ongoing CPR and keep the victim as
motionless as possible.
DON’T TOUCH PATIENT.
PRESS FLASHING SHOCK
BUTTON.
Warn all persons in attendance of the victim to stand clear and
stop touching the victim. Press the Shock Button to deliver
defibrillation therapy.
RELEASE SHOCK BUTTON.
Shock Button was depressed before the defibrillator was ready to
defibrillate. Release the Shock Button and press again after the
ready tone sounds.
SHOCK DELIVERED.
A defibrillation shock has just been delivered to the victim.
NO SHOCK DELIVERED.
No shock was delivered to the victim because rescuer failed to
press Shock Button or an error condition was detected.
n SHOCKS DELIVERED.
A total of n shocks have been delivered since the AED Plus was
turned on.
START CPR.
Begin CPR.
CONTINUE CPR.
Continue providing CPR. This prompt may also be issued if
Real CPR Help fails to detect chest compressions at least ¾ of an
inch deep.
PUSH HARDER.
CPR compressions are consistently less than 2 inches deep.
GOOD COMPRESSIONS.
After prompting to Push Harder, rescuer has succeeded in
delivering chest compressions at least 2 inches deep.
STOP CPR.
Stop CPR, the AED Plus is about to begin an ECG rhythm
analysis.
Voice Prompt
Definition