REV 2 RFD-200 S2 USER’S MANUAL

18

3.4

Using the RFD-200 S2’s Timer in Single Input Mode

The RFD-200 S2’s timer can also be started and stopped with only the

[TIMER START]

input.

This feature is handy for measuring the time duration of the

[TIMER START]

input when it

changes states and then returns to the previous state as shown below:

This features works only with the DRY or WET mode of the

[TIMER START]

input.

To select the single input mode, press and hold the

[TIMER STOP ARROW]

key for 2

seconds. The following screen will be displayed momentarily:

Use the steps below to measure the time a relay contact remains open:

a. Press and hold the

[TIMER STOP ARROW]

key for 2 seconds to select Single Input

Mode.

b. Connect the

[TIMER START]

inputs to the relay’s (Normally Closed) contact.

c. Select DRY mode for the

[TIMER START]

input.

d. Select START-ON from the control mode area.

e. Energize the relay coil to open the relay’s Normally Closed contact. The RFD-200 S2’s

timer will start and the time will be displayed on the LCD screen.

f. De-energize the relay coil to close the relay contact. The RFD-200 S2’s timer will stop

and the time will be displayed on the LCD screen.

Trigger Mode Set

To Single Input