REV 2 RFD-200 S2 USER’S MANUAL
10
2.4
Timer Start and Timer Stop Inputs
The RFD-200 S2’s built-in timer displays the test results in both
milliseconds and cycles.
There are three selections for the
[TIMER START]
input:
• INTERNAL
Timer starts when voltage/current sources are turned
on or off (user selects Start-On mode or Start-Off
mode).
• DRY
A change in state of a dry contact input after the user
has selected the Start-On mode or Start-Off mode will
start the timer.
• WET
A change in state of a wet contact input after the user has selected the Start-On mode
or Start-Off mode will start the timer.
There are three selections for the
[TIMER STOP]
input:
• INTERNAL
Timer stops when the user selects OFF mode.
• DRY
A change in state of a dry contact input after the timer has started will stop the timer.
• WET
A change in state of a wet contact input after the timer has started will stop the timer.
NOTES
1. A voltage level from 20 to 255 Vac/dc is considered as “ON”. Voltage less than 10
Vac/dc is considered as “OFF”. The input is not polarity sensitive.
2. The dry contact input detects contact closed (shorted) or contact opened.
3. Applying a voltage to the timer “Dry” contact input may damage the RFD-200 S2
2.5
RFD-200 S2 Power Resistors
Five power resistor connections are available on the RFD-200
S2. A finer control of the current output can be achieved by
connecting one or more resistors in series with the current
source.