AUTO-OHM 100/200 Series 2 Operating Procedures

30

Rev 3, 2009

APPENDIX A

Auto-Ohm Troubleshooting Guide

Item Symptom

Possible

Problem

Solution

1 Reading

is

incorrect.

1. Poor connection at
the test clamps.

1. Check connections to
ensure teeth of voltage-
sensing and current clamps
are firmly in contact with
the device under test.

2 “Cable

Error”

Message.

1. No test current
going through the
device under test.

2. Sensing cables
problem.

1. Check current cable
Connection to the device
under test.
2. Check sensing cable
connection.
3. Run Calibration Test.