beautypg.com

Performance verification, Required test equipment – Teledyne LeCroy ZD1500 User Manual

Page 19

background image

Instruction Manual

921921-00 Rev A

13

Performance Verification

This procedure can be used to verify the warranted characteristics of the ZD
Series High Impedance Active Probe.

The recommended calibration interval for the models ZD Series is one year. The
complete performance verification procedure should be performed as the first
step of annual calibration. Test results can be recorded on a photocopy of the
Test Record provided in Appendix A at the end of the manual.

Performance verification can be completed without removing the probe covers
or exposing the user to hazardous voltages. Adjustment should only be
attempted if a parameter measured in the Performance Verification Procedure is
outside the specification limits.

NOTE: Adjustment should only be performed by qualified personnel

This procedure tests the following specifications:

Output Zero Voltage

LF Attenuation Accuracy

Required Test Equipment

The following table lists the test equipment and accessories (or their
equivalents) that are required for performance verification of the ZD Series
Probes.

This procedure has been developed to minimize the number of calibrated test
instruments required.

Only the parameters listed in boldface in the Minimum requirements column
must be calibrated to the accuracy indicated.

Because the input and output connector types may vary on different brands and
models of test instruments, additional adapters or cables may be required.

This manual is related to the following products: