10. Troubleshooting


If you have a problem with your unit, please check the following list for solutions.

Symptom

Possible Cause / Solution

Refer to

The display does not appear
when the power is turned ON.

The AC adapter is not connected properly.

→ Confirm the AC adapter and power cable.

Low battery

→ Replace the batteries.
→ Recharge the batteries (Ni-MH batteries)

3.1

Ni-MH battery drains fast.

Insufficient battery charge → Recharge

3.1

Reading is blinking.

Alarm level is exceeded → Confirm alarm setting

4.3

Measurement does not start.

When operating status is:
WAIT → Wait until status changes to “READY”, and press

“START” key.

READY→ Press “START” key.
STOP → Press “START” key, wait until status changes to

“READY”, and press “START” key again.

4

Particle count or concentration
is too high.

Attach the supplied filter and confirm that the reading drops
to zero.

Reading drops to zero:
→ Concentration of measuring environment is too high.
Reading remains high:
→ Possible instrument failure.

Please contact your local distributor.


Particle count or concentration
is too low.

Confirm error status of laser power failure or flow error.

8

Reading is displayed as “##.#” “##.#” indicates that the measurable range is exceeded.


Printing cannot be performed.

- Setting such as the baud rate setting is not made properly.
→ Confirm the printer setting.
- Improper cable connection.

(Confirm that the printer cable is used. Not the RS232C
cable.)

6.4


Can not transfer data to the
computer.

- Confirm cable connection.

(Confirm that the RS232C cable is used. Not the printer

cable. )
- Computer is not properly set for data transfer.

6.3


If the problem cannot be solved by confirming the above, please contact your local distributor or
our service center.