4-6 ISO>4 Mode


Setup procedure is shown below.
(Character positions may not correspond to the actual screen.)


































Power Up

1.MEASURE MODE
2.DATA PROCESSING
3.OPTIONS

< MEASURE MODE >

1.REPEAT 4.CALC
2.SINGLE 5.REMOTE
3.CONT 6.ISO>4

Press ENTER key

ISO>4 0.3/0.5/5.0um
LOC.021 STR:N BEEP:N
SAMPLE 01:00 2TIMES
INT 00:10:00 OK

Press ENTER key

SAMPLING

SAMPLE TIME 00:01:00
TOTAL TIME 00:12:00
FREEREC.9999 OK

Press ENTER key

Press ENTER key

ISO>4 30B 20:32
WAIT 0.5 0.00E+0/m3

5.0 0.00E+0/m3

00/02 1POINT

Measurement Setting Screen

LOC.:

LOC number can be used for identifying room
numbers or measurements. Setting of this item
is not mandatory.

STR:

Used when storing data in the instrument. When
storing the data, select “Y” by using the ▲▼
key. Select “N” if you do not want to store the
data.

BEEP:

Alarm goes off when alarm level is exceeded.

SAMPLE:

Sets sampling time. Setting range is from 10 sec
to 99min 59sec.

TIMES:

Sets number of sampling times. Setting range is
from 1 to 99, or CONT (Repeats until storage
capacity is reached. Max. 10,000 times.)

INT:

Sets duration of one measurement cycle. Setting
range is from 10sec to 99min 59sec. Minimum
time is dependent on set sampling time.

Setting of every item is not mandatory.
For example, you can change only the LOC
number and press the START/STOP key, and the
display will switch to a measurement standby
screen.