9. Specifications


Product

Handheld Laser Particle Counter

Model

3887

Measuring Particle
Size

0.3, 0.5, 5.0μm
(Optional Specification: 0.3, 0.5, 1.0μm)

Flow Rate

0.1 cf/min (2.83 L/min)

Sampling Time

10 sec to 99min 59sec (1 sec increments)

Number of
Sampling Times

1 to 99 times or Continuous

Measurement
Modes

Total 6 modes:
Single / Repeat / Continuous / Calculation / ISO>4 / Remote

Display

LCD 20 letters, 4 lines

Error Display

Excess Concentration, Laser Power Failure, Flow Error (±10%), and
Low Battery

External I/O

USB Mini-B Connector (Wiring is different from USB)

Communication
Protocol

RS232C/RS485: Switched from the “Menu”.
RS232C is for communicating with a computer or printer.
RS485 is for cascade connection.
* In order to use RS485 for communicating with a computer, the computer must

be equipped with a RS485 I/F.

Baud Rate

9600bps (115200bps when linked by data transfer software.)

Buffer Memory

10,000 data records
(CALC mode requires 4 data records for 1 measurement.)

Power

Four (4) AA Ni-MH batteries (4.8V-12.1Ah), or AC adapter (input 100-240V)
*The batteries must be charged by the supplied battery charger. They cannot be

charged by the AC adapter.

Battery Life

Continuous operating hours: Max 3 hours when using Ni-MH batteries.

(Subject to change depending on operational conditions.)

Dimensions

111(W) × 70(H) × 197(D) mm

Weight

Approx. 680g (without battery)

Standard
Accessories

AC Adapter, Power Cable, Ni-MH Batteries, Battery Charger, Filter, Tube,
Communication Cable, Application Software, and Stand, Isokinetic Probe,
Traceability Certificate

Options

Printer, Printer cable, Tripod, Carrying Case

*Certain test functions required in China are not included