Caution – Gentec-EO P-LINK User Manual
Page 9

P-
LINK User’s Manual Revision 15.0
9
Gentec Electro
– Optics, Inc. All rights reserved
3
Front Panel Description :
See figure 3-1 below
1-
EXTERNAL POWER SUPPLY INPUT JACK:
Input voltage required: 9-12 VDC/100 mA.
Note: The external power supply input is provided only for the P-LINK with the RS-232
Serial Connection Option.
CAUTION
Permanent damage may occur to the optical meter if an external
power supply other than the GENTEC-EO 200130, 200960,
SPU15A-105 or SPU15A-104 is used. Please call GENTEC-EO or
your local distributor if extra power supplies are needed for a
particular setup.
2-
USB INTERFACE CONNECTOR:
This interface allows remote control and data transfers between the P-LINK and a computer
that has a USB communication port. With the RS-232 P-LINK, this connector is use as a RS-
232 com port.
3-
ANALOG OUTPUT:
For monitoring laser average power or energy by using external equipment such as a chart
recorder, a computer with an analog interface, a voltmeter, etc.
The output signal represents a DC analog voltage proportional to the amplified and anticipated
power detector response in the case of a power measurement. In the case of an energy
measurement (energy mode), the output signal is a DC voltage representing the pulse energy
value.
The user must enter the maximum value in the Settings-
>SET Max Analog Out Range …
menu. This value is the value at which the analog output equal 2.05 V. That provides the
best signal-to-noise ratio. The measured power or energy is then related to the output voltage
and to the selected range according to the following equations:
Vout = Measurement * 2.05 / Max Analog Out Range
For example with a 10W max analog range:
2.05 V corresponds to 10 W
1.025 V corresponds to 5 W
Another useful example: To set the analog output so that 1V corresponds to a measurement
of 56W, the Max Analog Out Range must be set to 20.5 according to the following equation:
Max Analog Out Range = 56 * 2.05
Specifications on the analog output:
Maximum output voltage:
2.05 V
Output impedance:
274
Connector type:
Female 1/8” jack
4-
PROBE INPUT JACK:
The P-LINK uses a DB-15 female connector to mate with the detector heads (probes).