beautypg.com

Gentec-EO P-LINK User Manual

Page 28

background image

P-

LINK User’s Manual Revision 15.0

28

Gentec Electro

– Optics, Inc. All rights reserved

*TLC

Modifies the trigger level in energy mode (+ 8 characters)

Example : *TLC2.00E-02 selects a Trig Level of 0.002 Joules

“ACK\r\r\n”

*NAM

Returns the model of the head to the serial port

“UP55N-400W-

H9\r\r\n

*VER

Returns the name of the software to the serial port

“Version 1.14\r\r\n”

*F01

Returns information about current status

All field are separate by a TAB character “\t”

“Version\t3\tName\tUP

55N-400W-

H9\tWavelength\t1064

\tTrig

Level\t2.500000e-

01\tMax Analog

Output\t2.047500e+00

\tMode\t0

Offset\t0\tMUL\t1.0000
00e+00\tOFF\t0.00000

0e+00\tPWCStatus

65536\tMinScale\t24\r\

r\n

*F02

Returns more information about current status

All field are separate by a TAB character “\t”

“Version\t3\tName

\tUP55N-400W-
H9\tWavelength

\t1064\tTrig

Level\t2.500000e-

01\tMax Analog

Output

\t2.047500e+00\tMode
\t0\tOffset\t0\tMUL\t1.0

0000e+00\tOFF\t0.000

000e+00\tPWCStatus\
t65536\tMinScale\t24\t
LinearCorr\t1\tAnticipa

tion\t1\tAttenuator

\t0\r\r\n

*MUL

Modifies the multiplication factor (+ 8 characters)

Example : *MUL1.00E+01 selects a multiplication factor of 10

“ACK\r\r\n”

*OFF

Modifies the offset (8 characters)

Example : *OFF-2. 0E-00 selects a user offset of two watts

“ACK\r\r\n”

*CMW

Turns the Energy Mode ON

“ACK\r\r\n”

*CMX

Turns the Energy Mode OFF. OFF by default.

“ACK\r\r\n”

*SHI

Make a physical scale down.

“ACK\r\r\n”

*SLO

Make a physical scale up.

“ACK\r\r\n”