Gentec-EO P-LINK User Manual
Page 28

P-
LINK User’s Manual Revision 15.0
28
Gentec Electro
– Optics, Inc. All rights reserved
*TLC
Modifies the trigger level in energy mode (+ 8 characters)
Example : *TLC2.00E-02 selects a Trig Level of 0.002 Joules
“ACK\r\r\n”
*NAM
Returns the model of the head to the serial port
“UP55N-400W-
H9\r\r\n
”
*VER
Returns the name of the software to the serial port
“Version 1.14\r\r\n”
*F01
Returns information about current status
All field are separate by a TAB character “\t”
“Version\t3\tName\tUP
55N-400W-
H9\tWavelength\t1064
\tTrig
Level\t2.500000e-
01\tMax Analog
Output\t2.047500e+00
\tMode\t0
Offset\t0\tMUL\t1.0000
00e+00\tOFF\t0.00000
0e+00\tPWCStatus
65536\tMinScale\t24\r\
r\n
”
*F02
Returns more information about current status
All field are separate by a TAB character “\t”
“Version\t3\tName
\tUP55N-400W-
H9\tWavelength
\t1064\tTrig
Level\t2.500000e-
01\tMax Analog
Output
\t2.047500e+00\tMode
\t0\tOffset\t0\tMUL\t1.0
0000e+00\tOFF\t0.000
000e+00\tPWCStatus\
t65536\tMinScale\t24\t
LinearCorr\t1\tAnticipa
tion\t1\tAttenuator
\t0\r\r\n
”
*MUL
Modifies the multiplication factor (+ 8 characters)
Example : *MUL1.00E+01 selects a multiplication factor of 10
“ACK\r\r\n”
*OFF
Modifies the offset (8 characters)
Example : *OFF-2. 0E-00 selects a user offset of two watts
“ACK\r\r\n”
*CMW
Turns the Energy Mode ON
“ACK\r\r\n”
*CMX
Turns the Energy Mode OFF. OFF by default.
“ACK\r\r\n”
*SHI
Make a physical scale down.
“ACK\r\r\n”
*SLO
Make a physical scale up.
“ACK\r\r\n”