Appendix IV: Torque Values Chart
All Values are Pounds (LB).
SIZE
GRADE 2
GRADE 2
GRADE 5
GRADE 5
GRADE 8
GRADE 8
LUBED
DRY
LUBED
DRY
LUBED
DRY
1/4-20
49 in
65 in
75 in
100 in
107 in
143 in
5/16-18
101 in
134 in
157 in
210 in
220 in
305 in
3/8-16
15 FT
20 FT
23 FT
31 FT
32.5 FT
44 FT
7/16-14
24 FT
30 FT
37 FT
50 FT
53 FT
70 FT
1/2-13
36.5 FT
49 FT
57 FT
75 FT
80 FT
107 FT
9/16-12
53 FT
70 FT
82 FT
109 FT
115 FT
154 FT
5/8-11
73 FT
97 FT
113 FT
151 FT
159 FT
211 FT
3/4-10
129 FT
173 FT
200 FT
266 FT
282 FT
376 FT
7/8-9
125 FT
166 FT
321 FT
430 FT
454 FT
606 FT
1-8
187.5 FT
250 FT
482.5 FT
640 FT
680 FT
900 FT
Appendix V: Tolerances:
Load Cell Base Plates, level within 1/64”" per Foot
Upper bearing Plates, level within 1/64”" per Foot
Check Stands, level within 1/64”" per Foot
Check Rods, level within 1/100" per Foot
Appendix VI: Load Cell Specifications:
Capacity:
200,000 LB.
Height:
7 1/4" Inches High.
Input Resistance:
450.50 Ohm
Output Resistance:
482.64 Ohm
Isolation Resistance:
>=5000 Meg Ohm (measured with 50 volts DC)
Calibration:
2.0 mV/V
Cable Length:
15 ft.
Wiring:
Excitation (+) = Green
Excitation (-) = Black
Output (+) = White
Output (-) = Red
Shield = Transparent