Section 3: Service & Maintenance

04/10

22 51191 Rev. 2

3.2.4. Load Cell Replacement, Continued

4a. For a Stainless Steel Scale, remove the six (6) Phillips-head screws securing

the Junction Box Cover.

4b. Open the Junction Box Cover, identifying the correct cable.

5. Cut the five (5) Load Cell Cable bundled

connections inside the Junction Box.

6a. If the Load Cell being replaced is in the

farthest position

, tie a drag string

to the cable, securing it with tape.

─ The cable for the

nearest position

Load Cells can be accessed and
pulled through much easier.

6b. Carefully pull the Load Cell Cable out

through the access hole, down the
Side Rail, tugging gently so the
string will stay fastened to it.

7. Carefully remove the replacement Load

Cell from its packing, inspecting it
thoroughly for any obvious shipping or
handling damages.

8. For Load Cell being replaced is in the

farthest position

, tie the drag string

to the

new

cable, securing it with

tape.

9. Pull the new Cable back trough the

Scale Platform, through the Pillar Access
Hole, and then into the Junction Box (if
Stainless Steel Scale
), using the
identical path as the old cable.

10. Place the replacement Load Cell onto its

Mounting Block, then secure it with the
two (2) Mounting Bolts.

11. Assemble foot to the new load cell,

tightening foot fully.

12. Torque each Mounting Bolt to 90 ft/lbs.

Be

very careful

to select the correct Load

Cell Cable before cutting and replacing it.

The Junction Box attaches with Velcro

®

.