9. TROUBLESHOOTING

If the communication is unavailable, check the following items.

• Whether all devices related to communication are turned on.

• Whether connections are correct.

• Whether the number of connected instruments and connection distance are as

specified

• Whether communication conditions coincide between the master station (host

computer) and slave stations (instrument)

• Transmission speed : 9600bps

• Data length : 8 bits

• Stop bit : 1 bit

• Parity : None

• Whether send/receive signal timing conforms to Section 5.6 in this manual.

• Whether the station No. designated as send destination by the master station coincides

with the station No. of the connected instrument.

• Whether more than one instrument connected on the same transmission line shares

the same station No.

• Whether the station No. of instruments is set at other than 0. If it’s 0, the

communication function does not work.