6.4 Read-out of read only word data [Function code:10
H
]
Function
code
Max word number write-
in in one message
Relative data
address
Resister No.
Kind of data
10
H
64 word
0000
H
006D
H
40001 40110 User setting
(1) Message composition
Command message composition
(byte)
Response message composition (byte)
Station No.
Station No.
Function code
Function code
Upper
Upper
Write-in start
No. (relative address)
Lower
Write-in designate
No. (relative address)
Lower
Upper
Upper
Write-in word number
Lower
Write-in word number
Lower
}
1 to 64
Lower
Write-in byte number
}
Write-in
word number
CRC data
Upper
Upper
First write-in word data
Lower
Upper
Next write-in word data
Lower
Upper
Last write-in word data
Lower
Lower
CRC Data
Upper
* Arrangement of write-in word data
MSB
LSB
Upper byte of contents of the first word data
Lower byte of contents of the first word data
Upper byte of contents of the next word data
Lower byte of contents of the next word data
Upper byte of contents of the last word data
Lower byte of contents of the last word data