Teledyne Analytical Instruments

19

(3) Message transmission (example)

The following shows an example of reading out from No. 1 station the setting CH5
measurement concentration, decimal point position and measuring unit.

Relative address of CH5 measurement concentration setting: 000C

H

Data number: 03

H



Command message composition (byte)

Response message composition (byte)

Station No.

01

H

Station No.

01

H

Function code

04

H

Function code

03

H

Read-out byte number

04

H

Upper

00

H

Upper

00

H

Read-out start No.
(relative address)

Lower

0C

H

Contents of the
first word data

Lower

00

H

Upper

00

H

Upper

03

H

Read-out word
number

Lower

03

H

Contents of the
next word data

Lower

E8

H

Lower

70

H

Upper

00

H

CRC data

Upper

08

H

Contents of the
last word data

Lower

00

H

Lower

81

H

CRC data

Upper

0D

H


* Meaning of read-out data

First word data contents 04 0D

H

= 1200

Next word data contents 00 02

H

= 2 (decimal point position)

Latest word data contents 00 00D

H

= 0 (vol %)


In the above case, measurement concentration = 12.00 vol%



*For handling of the decimal point and unit, refer to Section 7.1.