Teledyne 7600 - Infrared gas analyzer - communication manual User Manual
Page 19

Teledyne Analytical Instruments
19
(3) Message transmission (example)
The following shows an example of reading out from No. 1 station the setting CH5
measurement concentration, decimal point position and measuring unit.
Relative address of CH5 measurement concentration setting: 000C
H
Data number: 03
H
Command message composition (byte)
Response message composition (byte)
Station No.
01
H
Station No.
01
H
Function code
04
H
Function code
03
H
Read-out byte number
04
H
Upper
00
H
Upper
00
H
Read-out start No.
(relative address)
Lower
0C
H
Contents of the
first word data
Lower
00
H
Upper
00
H
Upper
03
H
Read-out word
number
Lower
03
H
Contents of the
next word data
Lower
E8
H
Lower
70
H
Upper
00
H
CRC data
Upper
08
H
Contents of the
last word data
Lower
00
H
Lower
81
H
CRC data
Upper
0D
H
* Meaning of read-out data
First word data contents 04 0D
H
= 1200
Next word data contents 00 02
H
= 2 (decimal point position)
Latest word data contents 00 00D
H
= 0 (vol %)
In the above case, measurement concentration = 12.00 vol%
*For handling of the decimal point and unit, refer to Section 7.1.